Your browser doesn't support javascript.
loading
Assessment of Imaging Distortion in Magnetic Resonance Imaging for Stereotactic Radiosurgery: Through Phantom Study
Journal of the Korean Society of Magnetic Resonance in Medicine ; : 7-13, 2000.
Artigo em Coreano | WPRIM | ID: wpr-166434
ABSTRACT

PURPOSE:

To assess the distortion of MRI with the Leksell stereotactic radiosurgery system in variable pulse sequence and imaging plane through phantom study, to find most adequate imaging plane and pulse sequence for stereotactic radiosurgery system. MATERIALS AND

METHODS:

We made the phantoms for MRI and get images in variable conditions and analyzed the image distortion using image analysis program, and statistically using paired student t-test.

RESULTS:

The transeverse plane images had acceptable error ranges (less than 1.5mm) in all pulse sequence in both the analysis of fiducial marker in stereotactic G-frame and the phantom study. The coronal plane images had unacceptable large errors (more than 1.5mm) in the analysis of fiducial marker in the stereotactic G-frame, but had corrected small errors (less than 1.5mm) in the phantom study.

CONCLUSION:

We find from the phantom study that the present MR machines are adequate for stereotactic surgery system in frequently used pulse sequences, and imaging planes.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Imageamento por Ressonância Magnética / Radiocirurgia / Marcadores Fiduciais Limite: Humanos Idioma: Coreano Revista: Journal of the Korean Society of Magnetic Resonance in Medicine Ano de publicação: 2000 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Imageamento por Ressonância Magnética / Radiocirurgia / Marcadores Fiduciais Limite: Humanos Idioma: Coreano Revista: Journal of the Korean Society of Magnetic Resonance in Medicine Ano de publicação: 2000 Tipo de documento: Artigo