Quality Assessment of Ultrasonographic Equipment Using an ATS-539 Multipurpose Phantom
Journal of the Korean Radiological Society ; : 533-541, 2008.
Artigo em Coreano | WPRIM | ID: wpr-172782
ABSTRACT

PURPOSE:

To determine the rate of congruence and to standardize assessment of US (ultrasound) phantom images with the use of an ATS-539 multipurpose phantom for US equipment currently utilized in Korea MATERIALS AND

METHODS:

US phantom images were scanned with a 3.0-5.0 MHz convex transducer and were digitized by use of an analogue-digital converter. Members of a committee with consent evaluated the US phantom images from 108 types of ultrasound equipment. The dead zone, vertical and horizontal measurement, axial/lateral resolution, focal zone, sensitivity, functional resolution and gray scale/dynamic range were evaluated. Congruence or incongruence of ultrasound equipment was determined based on the results of dead zone, axial/lateral resolution and gray scale/dynamic range measurements. Other factors were evaluated for the possibility as criteria with the use of the Mann-Whitney U test and receiver operator characteristic (ROC) curve analysis.

RESULTS:

The dead zone, axial/lateral resolution and gray scale/dynamic range were 91.7%, 94.4% and 76.9%, respectively, for suitable US equipment. Considering all three factors, 78 types of ultrasound equipment were passed. The congruence rate of focal zone and functional resolution were 62.4% and 69.3% of the US equipment, respectively.

CONCLUSION:

Of the US equipment, 72.2% of the equipment was acceptable based on the dead zone, axial/lateral resolution, and gray scale/dynamic range measurements as determined with the use of an ATS-539 phantom. Focal zone and 8 mm-functional resolution can be useful as a standard in the assessment of a US phantom image.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Controle de Qualidade / Transdutores / Coreia (Geográfico) Tipo de estudo: Estudo prognóstico País/Região como assunto: Ásia Idioma: Coreano Revista: Journal of the Korean Radiological Society Ano de publicação: 2008 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Controle de Qualidade / Transdutores / Coreia (Geográfico) Tipo de estudo: Estudo prognóstico País/Região como assunto: Ásia Idioma: Coreano Revista: Journal of the Korean Radiological Society Ano de publicação: 2008 Tipo de documento: Artigo