Your browser doesn't support javascript.
loading
The Efficacy of Metallic Stent Placement in the Treatment of Colorectal Obstruction
Korean Journal of Radiology ; : 79-86, 2002.
Artigo em Inglês | WPRIM | ID: wpr-180099
ABSTRACT
DBJECTIVE To evaluate the efficacy of newly designed covered and non-covered coated colorectal stents for colonic decompression. MATERIALS AND

METHODS:

Twenty-six patients, (15 palliative cases and 11 preoperative) underwent treatment for the relief of colorectal obstruction using metallic stents positioned under fluoroscopic guidance. In 24 of the 26, primary colorectal carcinoma was diagnosed, and in the remaining two, recurrent colorectal carcinoma. Twenty-one patients were randomly selected to receive either a type A or type B stent; for the remaining five, type C was used. Type A, an uncovered nitinol wire stent, was lightly coated to ensure structural integrity. Type B (flare type) and C (shoulder type) stents were polyurethane covered and their diameter was 24 and 26mm, respectively. The rates of technical success, clinical success, and complications were analyzed using the chi-square test, and to analyse the mean period of patency, the Kaplan-Meier method was used.

RESULTS:

Thirty of 31 attempted placements in 26 patients were successful, with a technical success rate of 96.8% (30/31) and a clinical success rate of 80.0% (24/30). After clinically successful stent placement, bowel decompression occurred within 1-4 (mean, 1.58+/-0.9) days. Five of six clinical failures involved stent migration and one stent did not expand after successful placement. In the preoperative group, 11 stents, one of which migrated, were placed in ten patients, in all of whom bowel preparation was successful. In the palliative group, 19 stents were placed in 15 patients. The mean period of patency was 96.25+/-105.12 days 146.25+/-112.93 for type-A, 78.82+/-112.26 for type-B, and 94.25+/-84.21 for type-C. Complications associated with this procedure were migration (n=6, 20%), pain (n=4, 13.3%), minor bleeding (n=5, 16.7%), incomplete expansion (n=1, 3.3%), and tumor ingrowth (n=1, 3.3%). The migration rate was significantly higher in the type-B group than in other groups (p=0.038).

CONCLUSION:

Newly designed covered and non-covered metallic stents of a larger diameter are effective for the treatment of colorectal obstruction. The migration rate of covered stents with flaring is higher than that of other types. For evaluation of the ideal stent configuration for the relief of colorectal obstruction, a clinical study involving a larger patient group is warranted.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Cuidados Paliativos / Doenças Retais / Neoplasias Colorretais / Stents / Doenças do Colo / Desenho de Equipamento / Obstrução Intestinal / Pessoa de Meia-Idade Tipo de estudo: Guia de Prática Clínica Limite: Adolescente / Adulto / Idoso / Aged80 / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2002 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Cuidados Paliativos / Doenças Retais / Neoplasias Colorretais / Stents / Doenças do Colo / Desenho de Equipamento / Obstrução Intestinal / Pessoa de Meia-Idade Tipo de estudo: Guia de Prática Clínica Limite: Adolescente / Adulto / Idoso / Aged80 / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2002 Tipo de documento: Artigo