Your browser doesn't support javascript.
loading
Comparison between the Set and the Measured Tidal Volume in the Pediatric Patients Using the Adult Volume-Limited Ventilator / 대한마취과학회지
Article em Ko | WPRIM | ID: wpr-18153
Biblioteca responsável: WPRO
ABSTRACT
In the volume-cycled ventilator, the actual tidal volume delivered to the patient is influenced by the compression volume and elasticity of the circuit. The purpose of the present study is to compare the set tidal volume with the measured tidal volume, and calculate the compression factor of the ventilator. We studied twenty pediatric patients weighting above 10 kg. The set tidal volume ( V(r)), the exhaled volume displayed in the ventilator ( V(exh)) and the actual tidal volume measured by the Wright spirometer ( V(sp)) were compared. The results were as follows: 1) Mean tidal volume was 248+/-92 ml, mean exhaled volume was 233+/-102 ml and mean spirometer volume was 19+/-97 ml. Thus the set and the measured tidal volume were different significantly (p<0.05). but we may calculate the actual tidal volume by the relationship with the set tidal volume ( Vsp-1.03V(T) - 56 ) 2) Mean compression factor was 1.35+/-0.92 ml/cmH2O.
Assuntos
Palavras-chave
Texto completo: 1 Índice: WPRIM Assunto principal: Ventiladores Mecânicos / Volume de Ventilação Pulmonar / Elasticidade Limite: Adult / Humans Idioma: Ko Revista: Korean Journal of Anesthesiology Ano de publicação: 1995 Tipo de documento: Article
Texto completo: 1 Índice: WPRIM Assunto principal: Ventiladores Mecânicos / Volume de Ventilação Pulmonar / Elasticidade Limite: Adult / Humans Idioma: Ko Revista: Korean Journal of Anesthesiology Ano de publicação: 1995 Tipo de documento: Article