Your browser doesn't support javascript.
loading
Development and Performance Evaluation of an Animal SPECT System Using Philips ARGUS Gamma Camera and Pinhole Collimator / 대한핵의학회잡지
Korean Journal of Nuclear Medicine ; : 445-455, 2005.
Artigo em Coreano | WPRIM | ID: wpr-200015
ABSTRACT

PURPOSE:

We developed an animal SPECT system using clinical Philips ARGUS scintillation camera and pinhole collimator with specially manufactured small apertures. In this study, we evaluated the physical characteristics of this system and biological feasibility for animal experiments. MATERIALS AND

METHODS:

Rotating station for small animals using a step motor and operating software were developed. Pinhole inserts with small apertures (diameter of 0.5, 1.0, and 2.0 mm) were manufactured and physical parameters including planar spatial resolution and sensitivity and reconstructed resolution were measured for some apertures. In order to measure the size of the usable field of view according to the distance from the focal point, manufactured multiple line sources separated with the same distance were scanned and numbers of lines within the field of view were counted. Using a Tc-99m line source with 0.5 mm diameter and 12 mm length placed in the exact center of field of view, planar spatial resolution according to the distance was measured. Calibration factor to obtain FWHM values in 'mm' unit was calculated from the planar image of two separated line sources. Tc-99m point source with 1 mm diameter was used for the measurement of system sensitivity. In addition, SPECT data of micro phantom with cold and hot line inserts and rat brain after intravenous injection of [I-123]FP-CIT were acquired and reconstructed using filtered back projection reconstruction algorithm for pinhole collimator.

RESULTS:

Size of usable field of view was proportional to the distance from the focal point and their relationship could be fitted into a linear equation (y=1.4x+0.5, x distance). System sensitivity and planar spatial resolution at 3 cm measured using 1.0 mm aperture was 71 cps/MBq and 1.24 mm, respectively. In the SPECT image of rat brain with [I-123]FP-CIT acquired using 1.0 mm aperture, the distribution of dopamine transporter in the striatum was well identified in each hemisphere.

CONCLUSION:

We verified that this new animal SPECT system with the Philips ARGUS scanner and small apertures had sufficient performance for small animal imaging.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Encéfalo / Calibragem / Câmaras gama / Tomografia Computadorizada de Emissão de Fóton Único / Experimentação Animal / Proteínas da Membrana Plasmática de Transporte de Dopamina / Injeções Intravenosas Tipo de estudo: Estudo prognóstico Limite: Animais Idioma: Coreano Revista: Korean Journal of Nuclear Medicine Ano de publicação: 2005 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Encéfalo / Calibragem / Câmaras gama / Tomografia Computadorizada de Emissão de Fóton Único / Experimentação Animal / Proteínas da Membrana Plasmática de Transporte de Dopamina / Injeções Intravenosas Tipo de estudo: Estudo prognóstico Limite: Animais Idioma: Coreano Revista: Korean Journal of Nuclear Medicine Ano de publicação: 2005 Tipo de documento: Artigo