Is Weight-Based Adjustment of Automatic Exposure Control Necessary for the Reduction of Chest CT Radiation Dose?
Korean j. radiol ; Korean j. radiol;: 46-53, 2010.
Article em En | WPRIM | ID: wpr-21024
Biblioteca responsável: WPRO
ABSTRACT
OBJECTIVE: To assess the effects of radiation dose reduction in the chest CT using a weight-based adjustment of the automatic exposure control (AEC) technique. MATERIALS AND METHODS: With Institutional Review Board Approval, 60 patients (mean age, 59.1 years; M:F = 35:25) and 57 weight-matched patients (mean age, 52.3 years, M:F = 25:32) were scanned using a weight-adjusted AEC and non-weight-adjusted AEC, respectively on a 64-slice multidetector CT with a 0.984:1 pitch, 0.5 second rotation time, 40 mm table feed/rotation, and 2.5 mm section thickness. Patients were categorized into 3 weight categories; 90 kg (n = 48). Patient weights, scanning parameters, CT dose index volumes (CTDIvol) and dose length product (DLP) were recorded, while effective dose (ED) was estimated. Image noise was measured in the descending thoracic aorta. Data were analyzed using a standard statistical package (SAS/STAT) (Version 9.1, SAS institute Inc, Cary, NC). RESULTS: Compared to the non-weight-adjusted AEC, the weight-adjusted AEC technique resulted in an average decrease of 29% in CTDIvol and a 27% effective dose reduction (p 91 kg weight groups, respectively, compared to 20.3, 27.9 and 32.8 mGy, with non-weight-adjusted AEC. No significant difference was observed for objective image noise between the chest CT acquired with the non-weight-adjusted (15.0 +/- 3.1) and weight-adjusted (16.1 +/- 5.6) AEC techniques (p > 0.05). CONCLUSION: The results of this study suggest that AEC should be tailored according to patient weight. Without weight-based adjustment of AEC, patients are exposed to a 17 - 43% higher radiation-dose from a chest CT.
Assuntos
Palavras-chave
Texto completo: 1 Índice: WPRIM Assunto principal: Doses de Radiação / Peso Corporal / Radiografia Torácica / Tomografia Computadorizada por Raios X Limite: Adult / Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Korean j. radiol Ano de publicação: 2010 Tipo de documento: Article
Texto completo: 1 Índice: WPRIM Assunto principal: Doses de Radiação / Peso Corporal / Radiografia Torácica / Tomografia Computadorizada por Raios X Limite: Adult / Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Korean j. radiol Ano de publicação: 2010 Tipo de documento: Article