Automated Lung Nodule Detection at Low-Dose CT: Preliminary Experience
Korean Journal of Radiology
;
: 211-216, 2003.
Artigo
em Inglês
| WPRIM
| ID: wpr-214909
ABSTRACT
OBJECTIVE:
To determine the usefulness of a computer-aided diagnosis (CAD) system for the automated detection of lung nodules at low-dose CT. MATERIALS ANDMETHODS:
A CAD system developed for detecting lung nodules was used to process the data provided by 50 consecutive low-dose CT scans. The results of an initial report, a second look review by two chest radiologists, and those obtained by the CAD system were compared, and by reviewing all of these, a gold standard was established.RESULTS:
By applying the gold standard, a total of 52 nodules were identified (26 with a diameter 5 mm). Compared to an initial report, four additional nodules were detected by the CAD system. Three of these, identified only at CAD, formed part of the data used to derive the gold standard. For the detection of nodules > 5 mm in diameter, sensitivity was 77% for the initial report, 88% for the second look review, and 65% for the CAD system. There were 8.0+/-5.2 false-positive CAD results per CT study.CONCLUSION:
These preliminary results indicate that a CAD system may improve the detection of pulmonary nodules at low-dose CT.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Doses de Radiação
/
Tomografia Computadorizada por Raios X
/
Programas de Rastreamento
/
Diagnóstico por Computador
/
Neoplasias Pulmonares
Tipo de estudo:
Estudo diagnóstico
/
Estudo de rastreamento
Limite:
Adulto
/
Idoso
/
Aged80
/
Humanos
Idioma:
Inglês
Revista:
Korean Journal of Radiology
Ano de publicação:
2003
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS