Delayed Type III Endoleak Caused by Fabric Erosion after Endovascular Repair of an Abdominal Aortic Aneurysm
The Korean Journal of Thoracic and Cardiovascular Surgery ; : 291-294, 2017.
Artigo em Inglês | WPRIM | ID: wpr-217609
ABSTRACT
A 74-year-old patient presented with recurrent aneurysms in the infrarenal abdominal aorta and right common iliac artery 6 years after endovascular aortic repair using endografts in the same location. The patient underwent an aorto-bi-iliac replacement with removal of the stent graft. Two holes measuring 2 mm each were found in the removed graft, and they appeared to have been caused by wear from continuous friction between the endograft and the aortic wall.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Aorta / Aorta Abdominal / Prótese Vascular / Fricção / Aneurisma da Aorta Abdominal / Transplantes / Endoleak / Procedimentos Endovasculares / Artéria Ilíaca / Aneurisma Limite: Idoso / Humanos Idioma: Inglês Revista: The Korean Journal of Thoracic and Cardiovascular Surgery Ano de publicação: 2017 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Aorta / Aorta Abdominal / Prótese Vascular / Fricção / Aneurisma da Aorta Abdominal / Transplantes / Endoleak / Procedimentos Endovasculares / Artéria Ilíaca / Aneurisma Limite: Idoso / Humanos Idioma: Inglês Revista: The Korean Journal of Thoracic and Cardiovascular Surgery Ano de publicação: 2017 Tipo de documento: Artigo