Your browser doesn't support javascript.
loading
Radiation Dose Reduction via Sinogram Affirmed Iterative Reconstruction and Automatic Tube Voltage Modulation (CARE kV) in Abdominal CT
Korean Journal of Radiology ; : 886-893, 2013.
Artigo em Inglês | WPRIM | ID: wpr-219663
ABSTRACT

OBJECTIVE:

To evaluate the feasibility of sinogram-affirmed iterative reconstruction (SAFIRE) and automated kV modulation (CARE kV) in reducing radiation dose without increasing image noise for abdominal CT examination. MATERIALS AND

METHODS:

This retrospective study included 77 patients who received CT imaging with an application of CARE kV with or without SAFIRE and who had comparable previous CT images obtained without CARE kV or SAFIRE, using the standard dose (i.e., reference mAs of 240) on an identical CT scanner and reconstructed with filtered back projection (FBP) within 1 year. Patients were divided into two groups group A (33 patients, CT scanned with CARE kV); and group B (44 patients, scanned after reducing the reference mAs from 240 to 170 and applying both CARE kV and SAFIRE). CT number, image noise for four organs and radiation dose were compared among the two groups.

RESULTS:

Image noise increased after CARE kV application (p or = 0.108). Effective doses decreased by 19.4% and 41.3% for groups A and B, respectively (all, p < 0.001) after application of CARE kV with or without SAFIRE.

CONCLUSION:

Combining CARE kV, reduction of mAs from 240 to 170 mAs and noise reduction by applying SAFIRE strength 3 or 4 reduced the radiation dose by 41.3% without increasing image noise compared with the standard-dose FBP images.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Doses de Radiação / Lesões por Radiação / Algoritmos / Radiografia Abdominal / Interpretação de Imagem Radiográfica Assistida por Computador / Estudos Retrospectivos / Desenho de Equipamento / Tomografia Computadorizada Multidetectores Tipo de estudo: Estudo observacional Limite: Adulto / Idoso / Aged80 / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2013 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Doses de Radiação / Lesões por Radiação / Algoritmos / Radiografia Abdominal / Interpretação de Imagem Radiográfica Assistida por Computador / Estudos Retrospectivos / Desenho de Equipamento / Tomografia Computadorizada Multidetectores Tipo de estudo: Estudo observacional Limite: Adulto / Idoso / Aged80 / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2013 Tipo de documento: Artigo