Microbial Contaminations of Ultrasonic Scalers, High-Speed Handpieces and Water Syringe Lines / 대한치주과학회지
The Journal of the Korean Academy of Periodontology
;
: 941-947, 1997.
Artigo
em Coreano
| WPRIM
| ID: wpr-22637
ABSTRACT
The purpose of this study was to evaluate various methods of decontamination of ultrasonic scalers, high-speed handpieces and air-water syringes in dental equipments. Eimination of possible sources of microbial infection in dental operatories should be of primary importance. Microbial contamination levels of high speed handpieces, air-water syringes and ultrasonic scalers of 11 dental units in Seoul National University Hospital were evaluated after flushing the lines, alcohol sponge rubbing, or soaking in 0.1% chlorhexidine for 1,3 and 5 minutes. The result suggests that flushing the lines or soaking the tips in 0.1% chlorhexidine before use of the water systems may reduce the microbial levels. Soaking in 0.1% chlorhexidine for 5 minutes was most effective in reducing bacterial contamination.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Poríferos
/
Seringas
/
Ultrassom
/
Clorexidina
/
Descontaminação
/
Equipamentos Odontológicos
/
Seul
/
Rubor
País/Região como assunto:
Ásia
Idioma:
Coreano
Revista:
The Journal of the Korean Academy of Periodontology
Ano de publicação:
1997
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS