Quality Control of Radiation Dosimetry Service / 의학물리
Korean Journal of Medical Physics
;
: 253-259, 2009.
Artigo
em Coreano
| WPRIM
| ID: wpr-227387
ABSTRACT
We have developed standards based on international criterions for the quality control of dose tested by the measurement institutions of individual exposure doses through improving the reliability of data on the exposure dose of individuals working in radioactive environment and securing the accuracy and reliability of individual dose measurements. Laws related to radiation dose applied to domestic institutions refer to ANSI N13.11?1993, but currently , in U.S. and some other countries the measurement of radiation doses is based on ANSI N13.11?2001 that reduced test categories and tightened the standards. We made efforts to simplify the standards and to reduce the number of dosimeters required in experiment, and avoided preventing or hindering the use of future technologies not approved under the current law such as glass dosimeter and optical stimulation dosimeter. The Quality Management Manual of Radiation Dosimetry Service, Assessment Manual of Radiation Dosimetry Service Accreditation Program, and the Personnel Dosimetry Performance-Criteria for Testing are documents applicable in supervising laboratories.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Controle de Qualidade
/
Radiometria
/
Vidro
/
Acreditação
/
Jurisprudência
Tipo de estudo:
Estudo prognóstico
Idioma:
Coreano
Revista:
Korean Journal of Medical Physics
Ano de publicação:
2009
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS