Your browser doesn't support javascript.
loading
A Case of Riata(R) Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
Korean Circulation Journal ; : 336-339, 2013.
Artigo em Inglês | WPRIM | ID: wpr-227417
ABSTRACT
A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventricular fibrillation. Chest radiography revealed a significant split in the insulation of the lead allowing the inner wire to protrude. We considered the removal of the failed lead, but the removal of ICD lead is potentially a high risk procedure, so we cut and capped a proximal part of the failed lead and inserted a new lead. This is the first report of a St. Jude Riata(R) dual coil defibrillator lead failure with clinical and radiologic evidence of a defect in lead insulation in Korea.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Pacientes Ambulatoriais / Choque / Estresse Psicológico / Tórax / Fibrilação Ventricular / Desfibriladores Implantáveis / Desfibriladores / Falha de Equipamento / Coreia (Geográfico) / Ruído Limite: Humanos País/Região como assunto: Ásia Idioma: Inglês Revista: Korean Circulation Journal Ano de publicação: 2013 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Pacientes Ambulatoriais / Choque / Estresse Psicológico / Tórax / Fibrilação Ventricular / Desfibriladores Implantáveis / Desfibriladores / Falha de Equipamento / Coreia (Geográfico) / Ruído Limite: Humanos País/Região como assunto: Ásia Idioma: Inglês Revista: Korean Circulation Journal Ano de publicação: 2013 Tipo de documento: Artigo