Anatomical Basis of Pronator Teres for Electromyography Needle Placement Using Ultrasonography
Annals of Rehabilitation Medicine
;
: 39-46, 2015.
Artigo
em Inglês
| WPRIM
| ID: wpr-22996
ABSTRACT
OBJECTIVE:
To find the optimal needle insertion site for needle electromyography of the pronator teres (PT) muscle among commonly used sites.METHODS:
Fifty forearms of 25 healthy subjects were evaluated. Four expected needle insertion points were designated as follows. Point 0 was positioned at the midpoint between the medial epicondyle and medial border of biceps tendon in the elbow crease. Points 1, 2, and 3 were located 2 cm, 3.5 cm and 5 cm distal to point 0, respectively. We assumed that the thickness of PT and the distances between a vertical line from each point to the medial margin of the PT were significant parameters for finding the optimal site. Thus, we measured these parameters through ultrasonographic examination.RESULTS:
In men, the PT was thickest at point 2, and in women, at point 1. The distance between the expected needle insertion line and medial margin of PT was longest at point 1 in both men and women, and was statistically significant compared to points 2 and 3. Both men and women had neurovascular bundles located lateral to the expected needle insertion line.CONCLUSION:
The most appropriate and safe needle electromyographic insertional site for the PT is 2-3.5 cm distal to the mid-point between the biceps tendon and medial epicondyle in the elbow crease and the needle should be inserted upward and medial.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Tendões
/
Ultrassonografia
/
Cotovelo
/
Eletromiografia
/
Antebraço
/
Agulhas
Tipo de estudo:
Estudo diagnóstico
Limite:
Feminino
/
Humanos
/
Masculino
Idioma:
Inglês
Revista:
Annals of Rehabilitation Medicine
Ano de publicação:
2015
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS