Your browser doesn't support javascript.
loading
Effect of stimulating pulse width on the threshold of electrically evoked compound action potential / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 1187-1190, 2014.
Artigo em Chinês | WPRIM | ID: wpr-234433
ABSTRACT
This paper discusses the relationship between stimulating pulse width and the threshold of electrically evoked compound action potential (ECAP). Firstly, the rheobase and chronaxy from strength-duration curve of nerve fiber was computed using the shepherd's experiment results. Secondly, based on the relationship between ECAP and the action potential of nerve fiber, a mathematical expression to describe the relationship between stimulating pulse width and ECAP threshold was proposed. Thirdly, the parameters were obtained and the feasibility was proved to the expression with the results of experiment using guinea pigs. Research result showed that with ECAP compared to the action potential of nerve fiber, their threshold function relationship with stimulating pulse width was similar, and rheobase from the former was an order smaller in the magnitude than the latter, but the chronaxy was close to each other. These findings may provide meaningful guidance to clinical ECAP measurement and studying speech processing strategies of cochlear implant.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Limiar Auditivo / Potenciais de Ação / Implantes Cocleares / Implante Coclear / Eletricidade / Potenciais Evocados / Cobaias / Condução Nervosa Tipo de estudo: Guia de Prática Clínica Limite: Animais Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Limiar Auditivo / Potenciais de Ação / Implantes Cocleares / Implante Coclear / Eletricidade / Potenciais Evocados / Cobaias / Condução Nervosa Tipo de estudo: Guia de Prática Clínica Limite: Animais Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2014 Tipo de documento: Artigo