A study of the effect of an anti-rotational inner post screw systems on abutment screw loosening for single implant : Part 1 / 대한치과보철학회지
The Journal of Korean Academy of Prosthodontics
;
: 345-353, 2007.
Artigo
em Inglês
| WPRIM
| ID: wpr-25861
ABSTRACT
STATEMENT OF PROBLEM:
Implant abutment screw joints tend to loosen under clinical conditions. Abutment screw loosening results in loss of preload in function.PURPOSE:
Anti-rotational inner post screw (ARIPS) systems were compared with conventional abutment screws to reduce screw loosening. Reverse torque values were evaluated. MATERIAL ANDMETHODS:
32 implant assemblies (Warentec, Co, Ltd, Seoul, Korea) were organized as the 30-Ncm-torque conventional groups and 30-Ncm-torque ARIPS groups in external and internal system. The specimens were tested to 10(6) cycles at a load of 200N. Preload reverse torque, postload reverse torque, and the ratio of postload reverse torque to preload reverse torque were evaluated. The data were analyzed with unpaired t-test in external and internal systems.RESULTS:
In the ratio of postload reverse torque to preload reverse torque, the ARIPS groups showed significant differences than the conventional screw group in both external and internal system.CONCLUSION:
Within the limitations of this study, abutment screw loosening was effectively reduced using ARIPS system.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Implantes Dentários
/
Torque
/
Seul
/
Articulações
País/Região como assunto:
Ásia
Idioma:
Inglês
Revista:
The Journal of Korean Academy of Prosthodontics
Ano de publicação:
2007
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS