Your browser doesn't support javascript.
loading
Discussion on testing of flow rate of infusion device about industry standard / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 292-294, 2014.
Artigo em Chinês | WPRIM | ID: wpr-259867
ABSTRACT
Carried on the exploration testing of flow rate of infusion device about industry standard YY 0451-2010. Engaged in flow rate experiments adopting different method that are provided by new and old industry standard for samples of the same type. Compared with the result of the dangerous coefficient by calculating the test data, the old standard can be more sensitive to reflect the situation of product flow rate, so it can be applied to conventional control of the products. The method which provided by the new industry standard is suitable for evaluating periodicity the level of product contaminated.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Bombas de Infusão / Fluxômetros Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS