Your browser doesn't support javascript.
loading
Analysis on DQA protocol of fMRI / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 1247-1250, 2010.
Artigo em Chinês | WPRIM | ID: wpr-260899
ABSTRACT
Our purpose is to introduce and analyze the data quality assurance (DQA) protocol of functional magnetic resonance imaging (fMRI). A water phantom was scanned to get DQA indexes. An fMRI sequence was used to get signal noise ratio (SNR) and Drift, which was calculated from maximum difference ratio of the average signal intensity in the region of interest (ROI) of image serials. The long period application of this method demonstrated that this DQA protocol can reflect imaging performance and the state of stability of the MRI scanner. Some application experience and discussion involved in DQA were also presented here.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Controle de Qualidade / Padrões de Referência / Algoritmos / Processamento de Imagem Assistida por Computador / Inteligência Artificial / Imageamento por Ressonância Magnética / Artefatos / Imagens de Fantasmas / Métodos Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2010 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Controle de Qualidade / Padrões de Referência / Algoritmos / Processamento de Imagem Assistida por Computador / Inteligência Artificial / Imageamento por Ressonância Magnética / Artefatos / Imagens de Fantasmas / Métodos Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2010 Tipo de documento: Artigo