The investigation of nickel dissolution from nitinol under different extraction conditions in biological tests / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 129-131, 2013.
Artigo em Chinês | WPRIM | ID: wpr-264254
ABSTRACT
The dissolution of nickel from nitinol alloy under different extraction conditions was investigated when biological tests of nitinol medical devices were carried out. It discussed the equivalence of these extraction conditions and found the best extraction conditions. In the experiment, two brands of nitinol were chosen and extracted under different extraction conditions according to GB/T 16886-12. Plasma emission spectrometer was used to analyse the concentration of nickel in extract liquid. The results show that there is no equivalence among these extraction conditions for nitinol materials. And it is suggested that (37 +/- 1) degree C, (72 +/- 2) h could be used as normal extraction condition and (70 +/- 2) degrees C, (24 +/- 2) h as accelerated extraction condition.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Temperatura / Titânio / Materiais Biocompatíveis / Teste de Materiais / Química / Ligas / Níquel Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2013 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS