Obstruction of the Scavenging Pathway during Surgery: A case report / 대한마취과학회지
Korean Journal of Anesthesiology ; : 778-780, 2007.
Artigo em Coreano | WPRIM | ID: wpr-26513
ABSTRACT
An acute obstruction of the breathing circuit can be devastating situation. An 56-year old man who was diagnosed distal femur fracture underwent open reduction with internal fixation. During surgery, an obstruction of the breathing circuit occurred and we found the cause of obstruction was the plugging of a gas disposal tube with ice and dust accumulation. After removing the ice and dust, ventilation normalized. We report this case of an obstruction of the scavenging pathway with a review of the relevant literature.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Respiração / Ventilação / Poeira / Fêmur / Gelo Limite: Humanos Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 2007 Tipo de documento: Artigo