Influence of Width and Thickness of DCI's Curved Section on Its Performance / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 10-12, 2016.
Artigo
em Chinês
| WPRIM
| ID: wpr-265585
ABSTRACT
By establishing a series of Dynamic Cervical Implants (DCI) within C5-C6 cervical spinal segments, the biomechanical finite element analysis for DCI with different width and thickness were carried out to investigate the influence of the width and thickness of DCI's curved section on its equivalent stress and range of motion (ROM), so as to provide some theoretical basis for the optimization of DCI's design. The results show that the width of DCI's curved section has more obvious influence on the ROM of lateral bending and torsion, in comparison with the thickness of DCI's curved section. By appropriate reduction in width, the ROMs of lateral bending and torsion increase obviously, i.e. the overall movement function of patients is improved. Furthermore, the increase of equivalent stress could be counteracted by corresponding increase of thickness.