Your browser doesn't support javascript.
loading
The Dose Effect of Isocenter Selection during IMRT Dose Verification with the 2D Chamber Array / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 222-224, 2015.
Artigo em Chinês | WPRIM | ID: wpr-265658
ABSTRACT
To investigate the dose effect of isocenter difference during IMRT dose verification with the 2D chamber array. The samples collected from 10 patients were respectively designed for IMRT plans, the isocenter of which was independently defined as P(o), P(x) and P(y). P(o) was fixed on the target center and the other points shifted 8cm from the target center in the orientation of x/y. The PTW729 was used for 2D dose verification in the 3 groups which beams of plans were set to 0 degrees. The γ-analysis passing rates for the whole plan and each beam were gotten using the different standards in the 3 groups, The results showed the mean passing rate of γ-analysis was highest in the P(o) group, and the mean passing rate of the whole plan was better than that of each beam. In addition, it became worse with the increase of dose leakage between the leaves in P(y) group. Therefore, the determination of isocenter has a visible effect for IMRT dose verification of the 2D chamber array, The isocenter of the planning design should be close to the geometric center of target.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Dosagem Radioterapêutica / Radioterapia de Intensidade Modulada / Raios gama / Métodos Limite: Humanos Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2015 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Dosagem Radioterapêutica / Radioterapia de Intensidade Modulada / Raios gama / Métodos Limite: Humanos Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2015 Tipo de documento: Artigo