Finite Element Analysis of Effect of Key Dimension of Nitinol Stent on Its Fatigue Behaviour / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 305-310, 2015.
Artigo em Chinês | WPRIM | ID: wpr-266681
ABSTRACT
To evaluate the fatigue behavior of nitinol stents, we used the finite element method to simulate the manufacture processes of nitinol stents, including expanding, annealing, crimping, and releasing procedure in applications of the clinical treatments. Meanwhile, we also studied the effect of the crown area dimension of stent on strain distribution. We then applied a fatigue diagram to investigate the fatigue characteristics of nitinol stents. The results showed that the maximum strain of all three stent structures, which had different crown area dimensions under vessel loads, located at the transition area between the crown and the strut, but comparable deformation appeared at the inner side of the crown area center. The cause, of these results was that the difference of the area moment of inertia determined by the crown dimension induced the difference of strain distribution in stent structure. Moreover, it can be drawn from the fatigue diagrams that the fatigue performance got the best result when the crown area dimension equaled to the intermediate value. The above results proved that the fatigue property of nitinol stent had a close relationship with the dimension of stent crown area, but there was no positive correlation.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Simulação por Computador / Stents / Análise de Falha de Equipamento / Análise de Elementos Finitos / Ligas Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2015 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Simulação por Computador / Stents / Análise de Falha de Equipamento / Análise de Elementos Finitos / Ligas Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2015 Tipo de documento: Artigo