Manual handling of defective channels in siemens somatom Sensation 16 CT / 生物医学工程学杂志
Journal of Biomedical Engineering
;
(6): 434-437, 2012.
Artigo
em Chinês
| WPRIM
| ID: wpr-271758
ABSTRACT
The present paper introduces the components and operating principle of Sensation 16 CT date measuring system (DMS), analyzes the fixed position of modules and channels in array detector modules (ADM), the Checkup table, Tuneup table and IRS table in ICS Syngo Service. The author also puts forward a solution of manual handling of defective channels.