Your browser doesn't support javascript.
loading
Comparison of fatigue fracture strength by fixture diameter of mini implants / 대한치과보철학회지
The Journal of Korean Academy of Prosthodontics ; : 156-161, 2012.
Artigo em Coreano | WPRIM | ID: wpr-27865
ABSTRACT

PURPOSE:

This study was conducted to obtain difference in fracture strength according to the diameter of one-body O-ring-type of mini implant fixture, to determine the resistance of mini implant to masticatory pressure, and to examine whether overdenture using O-ring type mini implant is clinically usable to maxillary and mandibular edentulous patients. MATERIALS AND

METHODS:

For this study, 13 mm long one body O-ring-type mini implants of different diameters (2.0 mm, 2.5 mm and 3.0 mm) (Dentis, Daegu, Korea) were prepared, 5 for each diameter. The sample was placed at 30degrees from the horizontal surface on the universal testing machine, and off-axis loading was applied until permanent deformation occurred and the load was taken as maximum compressive strength. The mean value of the 5 samples was calculated, and the compressive strength of implant fixture was compared according to diameter. In addition, we prepared 3 samples for each diameter, and applied loading equal to 80%, 60% and 40% of the compressive strength until fracture occurred. Then, we measured the cycle number on fracture and analyzed fatigue fracture for each diameter. Additionally, we measured the cycle number on fracture that occurred when a load of 43 N, which is the average masticatory force of complete denture, was applied. The difference on compressive strength between each group was tested statistically using one-way ANOVA test.

RESULTS:

Compressive strength according to the diameter of mini implant was 101.5 +/- 14.6 N, 149 +/- 6.1 N and 276.0 +/- 13.4 N, respectively, for diameters 2.0 mm, 2.5 mm and 3.0 mm. In the results of fatigue fracture test at 43 N, fracture did not occur until 2x106 cycles at diameter 2.0 mm, and until 5x106 cycles at 2.5 mm and 3.0 mm.

CONCLUSION:

Compressive strength increased significantly with increasing diameter of mini implant. In the results of fatigue fracture test conducted under the average masticatory force of complete denture, fracture did not occur at any of the three diameters. All of the three diameters are usable for supporting overdenture in maxillary and mandibular edentulous patients, but considering that the highest masticatory force of complete denture is 157 N, caution should be used in case diameter 2.0 mm or 2.5 mm is used.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Força de Mordida / Fraturas de Estresse / Colódio / Força Compressiva / Prótese Total / Revestimento de Dentadura / Fadiga Limite: Humanos Idioma: Coreano Revista: The Journal of Korean Academy of Prosthodontics Ano de publicação: 2012 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Força de Mordida / Fraturas de Estresse / Colódio / Força Compressiva / Prótese Total / Revestimento de Dentadura / Fadiga Limite: Humanos Idioma: Coreano Revista: The Journal of Korean Academy of Prosthodontics Ano de publicação: 2012 Tipo de documento: Artigo