Your browser doesn't support javascript.
loading
Hemocompatibility of polyoxymethylene used for bileaflet heart valve / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 347-351, 2014.
Artigo em Chinês | WPRIM | ID: wpr-290755
ABSTRACT
The possibility of polyoxymethylene (POM) as heart valve leaflet material was investigated by comparing the hemocompatibility with that of 316L stainless steel and low-temperature isotropic pyrolytic carbon (LTIC). Surface hydrophobicity was characterized by water contact angle measurement. Platelet adhesion, APTT/PT/TT and hemolysis rate tests were applied for evaluating hemocompatibility. The results showed that POM was hydrophobic and had a low hemolytic rate, adhesion amount and activation degree of platelets on POM surface were less than 316L stainless steel, and was similar to LTIC. This research pointed out potential application of POM as heart valve leaflets.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Resinas Sintéticas / Aço Inoxidável / Materiais Biocompatíveis / Plaquetas / Próteses Valvulares Cardíacas / Carbono / Química / Adesividade Plaquetária / Valvas Cardíacas Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Resinas Sintéticas / Aço Inoxidável / Materiais Biocompatíveis / Plaquetas / Próteses Valvulares Cardíacas / Carbono / Química / Adesividade Plaquetária / Valvas Cardíacas Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2014 Tipo de documento: Artigo