Your browser doesn't support javascript.
loading
A long term accelerating corrosion fatigue texting of coronary stents in vitro / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 398-401, 2008.
Artigo em Chinês | WPRIM | ID: wpr-291225
ABSTRACT
According to the related standards, an in vitro corrosion fatigue testing of coronary stents was designed. The stents were fixed in the latex tubes, which were full of 0.9% saline solution, and radial stress was produced for simulating natural vessel. The accelerated fatigue test was performed with 4 x 10(8) cycles at a frequency of 60 Hz, which was equal to 10 years in vivo implantation. Twelve coronary stents made from stainless steel were adopted in the experiment. The bulk structure and surface morphology before and after testing were analysed by scanning electron microscopy. The structure damage and surface change caused by corrosion fatigue were identified and the probable reasons were proposed.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Aço Inoxidável / Fatores de Tempo / Materiais Biocompatíveis / Simulação por Computador / Teste de Materiais / Stents / Química / Análise de Falha de Equipamento / Corrosão / Modelos Cardiovasculares Tipo de estudo: Estudo prognóstico Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2008 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Aço Inoxidável / Fatores de Tempo / Materiais Biocompatíveis / Simulação por Computador / Teste de Materiais / Stents / Química / Análise de Falha de Equipamento / Corrosão / Modelos Cardiovasculares Tipo de estudo: Estudo prognóstico Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2008 Tipo de documento: Artigo