We have carried out a research on the method for Computer Aided Design and Manufacturing (CAD/ CAM) of the framework of porcelain-fused-to-metal(PFM) fixed bridge. Taking the missing of the mandibular first molar as an example, firstly, we used ATOS II 3D scanner to collect the 3D data of the abutment teeth and used the acquried data to design the coping crowns. Secondly, the relevant standard bridge was selected from the standard bridges database and was deformed inaccordance to the characteristic of the patient. Then it was set to the correct position between the coping crowns automatically based on the restricting lines and faces. Finally, the connector was designed to connect the coping crowns and the bridge based on the touching area. The final product was manufactured by 3-axis NC milling machine. The results of examination showed that the framework of PFM fixed bridge fitted the abutment teeth very well, and the hardness of the connector and the precision of the whole model were feasible and reasonable. The method of designing and manufacturing the framework of PFM fixed bridge from CAD/CAM is practicable, efficient and accurate. It is also an important part of homemade dental CAD/CAM system.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Dente Suporte
/
Ligas Metalo-Cerâmicas
/
Planejamento de Prótese Dentária
/
Desenho Assistido por Computador
/
Coroas
/
Prótese Parcial Fixa
/
Métodos
/
Dente Molar
Tipo de estudo:
Estudo prognóstico
Limite:
Humanos
Idioma:
Chinês
Revista:
Journal of Biomedical Engineering
Ano de publicação:
2008
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS