Design and simulation of interference current therapeutic instruments based on FPGA / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 100-104, 2008.
Artigo em Chinês | WPRIM | ID: wpr-309634
ABSTRACT
This paper applies the DDFS theory, uses the method based on the Look-Up-Table, and designs an interference current therapeutic instrument with the automatic frequency scanning module and DDFS module to generate two different interference currents. The design of every module and the whole system are simulated by quartus II, the result shows that the design principle is correct and feasible, and worthy of applications.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Processamento de Sinais Assistido por Computador / Terapia por Estimulação Elétrica / Desenho de Equipamento Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2008 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Processamento de Sinais Assistido por Computador / Terapia por Estimulação Elétrica / Desenho de Equipamento Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2008 Tipo de documento: Artigo