Measurement of Serum Levels of 25-Hydroxyvitamin D3 and 25-Hydroxyvitamin D2 Using Diels-Alder Derivatization and Ultra-Performance Liquid Chromatography-Tandem Mass Spectrometry
Laboratory Medicine Online
;
: 188-196, 2012.
Artigo
em Coreano
| WPRIM
| ID: wpr-31226
ABSTRACT
BACKGROUND:
Vitamin D has been recently shown to play important roles in the functioning of various systems. Most of the current analytical methods for measuring vitamin D levels are based on immunoassays. We simultaneously measured the levels of 25-hydroxyvitamin D3 [ 25(OH)D3 ] and 25-hydroxyvitamin D2 [ 25(OH)D2 ] in human serum by performing ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) after Diels-Alder derivatization with 4-phenyl-1,2,4-triazoline-3,5-dione (PTAD) and evaluated the performance of our method.METHODS:
After liquid-liquid extraction, samples were dried under N2 at 50degrees C for 1 hr followed by Diels-Alder derivatization with ethyl acetate containing 0.1 mg/mL PTAD. The samples were resuspended in 60 microL of methanol10 mM ammonium formate solution (11, V/V). C18 UPLC column and positive ion multiple reaction monitoring transitions such as m/z 558.35-->298.1, 25(OH)D3; m/z 570.35-->298.1, 25(OH)D2; and m/z 564.35-->298.1, hexadeuterated-25(OH)D3 were used for UPLC-MS/MS.RESULTS:
The within-run imprecision (CVs) for 25(OH)D3 and 25(OH)D2 were 3.5-4.0% and 3.8-4.2%, respectively, and the corresponding between-run CVs were 3.3-5.5% and 4.7-5.8%. The lower limit of quantification for 25(OH)D3 and 25(OH)D2 were 0.5 and 1.0 ng/mL, respectively. The curve for interassay calibration variability data obtained over concentrations of 0-120 ng/mL for 25(OH)D3 and 0-90 ng/mL for 25(OH)D2 was linear and reproducible [ 25(OH)D3, R2=0.993; 25(OH)D2, R2=0.998]. The total 25(OH)D levels in Koreans (average, 18.7 ng/mL) were lower than those in American Caucasians, and the percentage of people with total 25(OH)D levels under 10 ng/mL was 8.1%.CONCLUSIONS:
Our method to measure 25(OH)D3 and 25(OH)D2 levels by performing UPLC-MS/MS after PTAD derivatization showed good performance as a sensitive and reproducible method for routine analysis of vitamin D status.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Espectrometria de Massas
/
Triazóis
/
Vitamina D
/
Calcifediol
/
Calibragem
/
Imunoensaio
/
25-Hidroxivitamina D 2
/
Espectrometria de Massas em Tandem
/
Extração Líquido-Líquido
/
Formiatos
Limite:
Humanos
Idioma:
Coreano
Revista:
Laboratory Medicine Online
Ano de publicação:
2012
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS