Correlation of examination couch position and height with image registration accuracy in an effective field of vision of SPECT/CT / 南方医科大学学报
Journal of Southern Medical University ; (12): 1181-1183, 2014.
Artigo em Chinês | WPRIM | ID: wpr-312612
ABSTRACT
<p><b>OBJECTIVE</b>To explore the correlation of the position and height of the examination couch with the accuracy of image registration in an exact effective field of vision (FOV) of single photon emission computed tomography/computed tomography (SPECT/CT) and determine the optimal position of the examination couch for best image registration.</p><p><b>METHODS</b>The tests were performed on a SPECT/CT with 6 radioactive (153)Gd sources in 3 groups (two in each group with energy peak of 40.9 Kev, 41.54 Kev, and 103.18 Kev) in an assorted standard pattern. Thirteen groups of slice data were obtained by changing the position and height of the bed but keeping the tomograpy slice constant. The bias of the accuracy of image registration was acquired using ALIGNMENT software and compared for different examination couch positions and heights.</p><p><b>RESULTS</b>The position of the examination couch at 0 cm in the Z axis and -4 cm in the Y axis showed the least error of image registration.</p><p><b>CONCLUSION</b>To achieve optimal SPECT and CT image fusion, quality control must be implemented and calibration of image registration accuracy should be done when necessary.</p>
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Algoritmos / Software / Aumento da Imagem / Tomografia Computadorizada de Emissão de Fóton Único / Posicionamento do Paciente Limite: Humanos Idioma: Chinês Revista: Journal of Southern Medical University Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Algoritmos / Software / Aumento da Imagem / Tomografia Computadorizada de Emissão de Fóton Único / Posicionamento do Paciente Limite: Humanos Idioma: Chinês Revista: Journal of Southern Medical University Ano de publicação: 2014 Tipo de documento: Artigo