Your browser doesn't support javascript.
loading
Determination of residual epoxyethane in medical instruments by headspace capillary gas chromatography / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 201-230, 2007.
Artigo em Chinês | WPRIM | ID: wpr-323285
ABSTRACT
<p><b>OBJECTIVE</b>To establish a method of determining the residual epoxyethane in medical instruments.</p><p><b>METHOD</b>An HP-5 capillary column (30 m x 0.32 mm x 0.25 microm), an FID detecter and an oven were used in the determination. The oven temperature was kept at 120 degrees C for 8 minutes, the example inlet temperature was kept at 200 degrees C, while the detecter temperature was kept at 250 degrees C.</p><p><b>RESULT</b>The method has a linear range between 5.1920-51.920 microg, (r = 0.997, n = 6). The average rate of recovery of the injecter is 103.3%. The average rate of recovery of the infusion devices is 81.04%.</p><p><b>CONCLUSION</b>The method has been proved to be convenient, sensitive and accurate. It is suitable for the determination of residual organic solvents in medical instruments.</p>
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Solventes / Cromatografia Gasosa / Equipamentos Descartáveis / Compostos de Epóxi / Etano / Métodos Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2007 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS