Comparison of three dynamic algorithms of electrical impedance tomography based on simulation study / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 4-7, 2012.
Artigo
em Chinês
| WPRIM
| ID: wpr-325949
ABSTRACT
Three algorithms of electrical impedance tomography (EIT) are studied in this paper. The image resolution, anti-noise property and computation rapidity of the reconstruction algorithms are compared. As a result, it shows that back-projection algorithm has good anti-noise property, that NOSER algorithm generates images with good resolution, and that sensitivity matrix algorithm has moderate property.