ABSTRACT
A non-invasive detecting and analyzing method which used to monitor the working condition of rotating anode X-ray was proposed. Based on the NI development environment, accelerometer, 24-bit high resolution data acquisition card and personal computer were connected to construct the system for collecting the vibration signal of X-ray tube. Results demonstrate that the system could acquire and store the vibration data of X-ray tube quickly and efficiently. The characteristics of vibration, were extracted and processed, which proposed a new approach to detect the malfunction of rotating anode X-ray early and effectively.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Vibração / Software / Tomografia Computadorizada por Raios X / Desenho de Equipamento / Métodos Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2011 Tipo de documento: Artigo