Your browser doesn't support javascript.
loading
Circulatory Support for the Operation of the Descending Thoracic Aorta with a Heparin-coated Tube and a Roller Pump / 日本心臓血管外科学会雑誌
Japanese Journal of Cardiovascular Surgery ; : 515-518, 1992.
Artigo em Japonês | WPRIM | ID: wpr-365853
ABSTRACT
We applied a new circulatory support system to three patients with descending aortic operation between 1989 and 1990. The new system consisted of a polyvinylchloride tube coated with hydrophilic heparinized polymer and a roller pump. Low doses of heparin (0.5mg/kg) were administered in two patients and none in one patient before aortic bypass. The aortic crossclamping time ranged between 52 and 64 minutes. In two patients, whose inflow cannulae were 18Fr. or 20Fr., successful bypass flow exceeded 2<i>l</i>/min and optimal distal aortic pressure were maintained during aortic cross-clamping. The postoperative courses of these patients were uneventful, free from thromboembolic episodes and subsequently discharged from hospital. Furthermore, postoperative careful examination revealed no thrombus formation occurred in the three devices. In conclusion, our new simple, antithrombogenic circulatory system is extremely useful for descending aortic operations.
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Japonês Revista: Japanese Journal of Cardiovascular Surgery Ano de publicação: 1992 Tipo de documento: Artigo