Quality control and quality assurance of single photon emission computed tomography/computed tomography ( SPECT/CT) system / 中国基层医药
Chinese Journal of Primary Medicine and Pharmacy
;
(12): 2600-2602, 2010.
Artigo
em Chinês
| WPRIM
| ID: wpr-386359
ABSTRACT
Objective To investigate the quality control and quality assurance of the SPECT/CT system.Methods The energy peak,energy resolution capacity and the intrinsic uniformity of the device's detector were regularly examined, and the quality control was performed. Results The daily hardware maintain could reduce the rate of system's trouble. The device's energy peak,energy resolution capacity were consistent during half year's observation period. The two detector's intrinsic uniformity were better after calibration than before. [ detector Ⅰ ( 2.71 ± 0.28 ) vs (2.37±0.11)(t=2.489,P<0. 05);detector Ⅱ(2.68 ±0.12)vs(2.38 ±0. 19)(t =6.421,P <0.01) ] .Conclusion Regular quality control and maintain could keep the function stabilization,enhance the availability rate of the SPECT/CT system.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Idioma:
Chinês
Revista:
Chinese Journal of Primary Medicine and Pharmacy
Ano de publicação:
2010
Tipo de documento:
Artigo