ABSTRACT

PURPOSE:

To evaluate the refractive predictability of a partial coherence interferometry (PCI) biometry device (IOL Master(R)) for cataract surgery and to investigate factors that may affect it.

METHODS:

Retrospective review of 209 eyes from 151 patients that had undergone preoperative PCI biometry and an uneventful phacoemulsification cataract surgery with posterior chamber intraocular lens (IOL) implantation was conducted. Prediction error defined as the intended refraction minus the postoperative refraction in spherical equivalent (SE) and the absolute error were analyzed according to IOL calculation formulas, patient characteristics, preoperative visual acuity (VA) and refraction, posterior subcapsular cataract (PSC), signal-to-noise ratio (SNR), and axial length (AL).

RESULTS:

The overall refractive predictability of the PCI device was good. Generally, the SRK/T formula performed better than the SRK-II formula. Refractive predictability was slightly worse in eyes with > or =+2.0 diopters (D) of preoperative SE (with both SRK-II and SRK/T) and in eyes with an AL or =+2.0D was related to a significantly greater hyperopic shift in postoperative refraction. With proper verification of measured data and a suitable IOL calculation formula, good refractive predictability is expected from PCI biometry regardless of patient characteristics, preoperative VA, SNR, PSC, and AL.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Refração Ocular / Erros de Refração / Acuidade Visual / Reprodutibilidade dos Testes / Estudos Retrospectivos / Facoemulsificação / Implante de Lente Intraocular / Interferometria / Lentes Intraoculares / Luz Tipo de estudo: Estudo observacional / Estudo prognóstico Limite: Idoso / Aged80 / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Ophthalmology Ano de publicação: 2009 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Refração Ocular / Erros de Refração / Acuidade Visual / Reprodutibilidade dos Testes / Estudos Retrospectivos / Facoemulsificação / Implante de Lente Intraocular / Interferometria / Lentes Intraoculares / Luz Tipo de estudo: Estudo observacional / Estudo prognóstico Limite: Idoso / Aged80 / Feminino / Humanos / Masculino Idioma: Inglês Revista: Korean Journal of Ophthalmology Ano de publicação: 2009 Tipo de documento: Artigo