Your browser doesn't support javascript.
loading
Continuous Detection Method for Ractopamine with Surface Plasmon Resonance Biosensor / 分析化学
Chinese Journal of Analytical Chemistry ; (12): 381-384, 2010.
Artigo em Chinês | WPRIM | ID: wpr-403203
ABSTRACT
The interaction between the anti-Ractopamine (Rac) monoclonal antibody and the Rac derivation immobilized on the sensor chip surface was studied with surface plasmon resonance (SPR) biosensor. A continuous detection method based on the linear response during the association phase was developed. The sensor chip surface was regenerated after several tests performed continuously, thus the detection step was simplified and the life span of the chip was prospected to be prolonged. The detection was performed as an inhibitive immunoassay. The mixture of anti-Rac monoclonal antibody and the sample flowed over the surface with Rac derivation was immobilized. The relative response was in inversely proportion to the concentration of Rac. The detection limit was less than 4 μg/L with a detection time of 15 min.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: Chinese Journal of Analytical Chemistry Ano de publicação: 2010 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: Chinese Journal of Analytical Chemistry Ano de publicação: 2010 Tipo de documento: Artigo