Your browser doesn't support javascript.
loading
Determination of Inorganic Mercury and Total Organomercury in ChemicalWaste Water by Cold Vapor Generation-Atomic Fluorescence Spectrometry / 分析化学
Chinese Journal of Analytical Chemistry ; (12): 403-405, 2001.
Artigo em Chinês | WPRIM | ID: wpr-410552
ABSTRACT
A sensitive method for the determination of inorganic mercury and total organomercury in chemical waste water by cold vapor generation-atomic fluorescence spectrometry (CVG-AFS) was proposed. Inorganic mercury was directly determined by CVG-AFS. Organomercury was oxidized by potassium persulfate to its inorganic form prior to analysis. The difference between the two data was the value of total organomercury. Parameters affecting the sensitity in the process for the determination of mercury had been studied and optimized. Detection limit of mercury chloride was 8.2 ng/L.
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Analytical Chemistry Ano de publicação: 2001 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Analytical Chemistry Ano de publicação: 2001 Tipo de documento: Artigo