Your browser doesn't support javascript.
loading
Study of rapid endotoxin detection based on interdigital gold electrode and SAW / 国际生物医学工程杂志
International Journal of Biomedical Engineering ; (6): 21-23, 2012.
Artigo em Chinês | WPRIM | ID: wpr-424882
ABSTRACT
ObjectiveTo establish a method of rapid detection of bacteria endotoxin based on interdigital gold electrode and surface acoustic wave (SAW).MethodsWith the carriers of 3-APTES and glutaral pentanedial,Polymyxin B (PMB) was immobilized on interdigital gold electrode,and then reacted with the bacteria endotoxin' s biological active material (lipopolysacchride,LPS) selectively.Double differential circuit based on SAW was employed to complete the detection.Two output oscillation signals were differentiated through mixer and the endotoxin content was described by frequency shift.ResultsThe experimental results indicated that there was a liner relationship between the frequency shift and endotoxin content within the content range of 75-130 EU/ml.ConclusionComparing with traditional method,the detection time is shortened to 10 minutes,the sensitivity is 13.8 KHz/(EU·mL-1),and repeatability is above 90%.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: International Journal of Biomedical Engineering Ano de publicação: 2012 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: International Journal of Biomedical Engineering Ano de publicação: 2012 Tipo de documento: Artigo