Analysis the effects of film dosimetry during processing and scanning / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology
;
(6): 285-288, 2012.
Artigo
em Chinês
| WPRIM
| ID: wpr-425928
ABSTRACT
ObjectiveTo explore the effects of different flushing and scanning processing on optical density (A) responding and set up the clinic quality assurance protocol based on silver halide emulsions radiographic films.MethodsSetting different flushing temperature and choosing different batch's film and developer and fixer or fixer in the same batch with different analyzing dose were performed to analyze the effect on A value; The effect of light uniform,the stabilize work time and the noise of different scanning resolution were discussed.ResultsThe A value at the same dose level would enhance as the temperature increased;the responding curve of dose and A value with different batches of films varied a lot;the responding curve of dose and A value with different batches of films and developer and fixer solution had marked variations;the responding of dose and A with the same batches would show some low.The heterogeneity of the scanner would achieve 0.03 ; the A value of the same dose would gradually steady while the time ofscanning was more than 10 minutesthe affect of noise would increase as the dose and resolution ratio increased.ConclusionsThe best processing temperature is 29-31 ℃.Different batches of film couldn't be confounded.A new calibration must be obtained when the film dosimetry to evaluate dose distribution is used.A 5 - 10 min warm up for stabilize work and the best setting resolution/depth lever are 72 dpi/16 bit for scanning films are determined.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Tipo de estudo:
Guia de Prática Clínica
Idioma:
Chinês
Revista:
Chinese Journal of Radiation Oncology
Ano de publicação:
2012
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS