A comparative study on a new transnasal gastroscopy and routine gastroscopy / 中华消化内镜杂志
Chinese Journal of Digestive Endoscopy
;
(12): 579-581, 2013.
Artigo
em Chinês
| WPRIM
| ID: wpr-439398
ABSTRACT
Objective To investigate the clinical value of new transnasal Fujifilm EG-530-NW electronic gastroscope.Methods A total of 295 patients who underwent gastroscopy were randomly divided into 2 groups to receive examination by routine endoscope (n =172) or Fujifilm EG-530-NW transnasal endoscope (n =123).The fluctuation of heart rate,blood pressure and compliance during the examination were recorded and compared between the two groups.Results The new Fujifilm EG-530-NW could obtain clear images with same quality of routine gastroscope.The heart rate change and visual analog pain score in patients from EG-530-NW gastroscope group were significantly lower than those of routine gastroscope group (P < 0.05).No significant difference in blood pressure fluctuation was observed between the two groups.In patients older than 50 years,the fluctuations of heart rate and blood pressure in EG-530-NW group were significantly less than those of routine gastroscope group.Conclusion Fujifilm EG-530-NW gastroscope can get high quality image with less influence on heart rate and blood pressure fluctuation,which is more significant in older patients.