Awareness during general anesthesia due to the unlock position of the vaporizer : A report on two cases / 대한마취과학회지
Korean Journal of Anesthesiology ; : 693-697, 2009.
Artigo em Coreano | WPRIM | ID: wpr-44232
ABSTRACT
Awareness with recall after general anesthesia is a relatively infrequent occurrence and this awareness is often associated with significant adverse psychological sequelae, including posttraumatic stress disorder. The occurrence of awareness is often the consequence of the light anesthetic techniques with using minor anesthetic agents. Because the locking lever of the Penlon Sigma Delta Anesthetic vaporizer (sevoflurane) was at the unlock position, disconnection happened between the anesthetic machine and the vaporizer (the Selectatec(R) vaporizing system). In our two cases, the inhalation agents were not supplied and awareness with recall was identified after surgery. We recommend to check the concentration of the inhalation agents by monitoring the concentration of using endtidal inhalation agents and to measure the hypnotic level of anesthesia via the bispectral index (BIS) and the auditory evoked potential (AEP).
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Transtornos de Estresse Pós-Traumáticos / Nebulizadores e Vaporizadores / Inalação / Potenciais Evocados Auditivos / Anestesia / Anestesia Geral / Anestésicos / Luz Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 2009 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Transtornos de Estresse Pós-Traumáticos / Nebulizadores e Vaporizadores / Inalação / Potenciais Evocados Auditivos / Anestesia / Anestesia Geral / Anestésicos / Luz Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 2009 Tipo de documento: Artigo