Your browser doesn't support javascript.
loading
Development and biomechanical evaluation of a model of lumbar interbody fusion device / 重庆医学
Chongqing Medicine ; (36): 465-468, 2014.
Artigo em Chinês | WPRIM | ID: wpr-445288
ABSTRACT
objective To introduce the design concept and structure of a new type of lumbar intervertebral fusion cage ,and to e-valuate its biomechanical properties .Method A partially bioasorbable interbody fusion cage(PBIFC) made of nano hydroxyapatite and poly amino acid /calcium sulfate copolymer materials was developed .Range of motion(ROM ) ,compressive load ,and pull-out test on flexion ,extension ,lateral bending ,and torsion moment on fresh calf L3/L4 specimens of functional spinal union were carried out of iliac bone group ,PBIFC group ,and nano hydroxyapatite/polyamide 66(nHA/PA66) group .Result Of each movement ,the ROM value of iliac bone group are higher than the other two groups ,the difference was statistically significant (P0 .05) .The pull-out strength of PB-IFC group are higher than iliac bone group ,and the difference was statistically significant (P0 .05) .The compressive load of iliac bone group was lower than that of two cage group ,the difference was statistically significant (P0 .05) .Conclusion With good implant stability ,pull-out resistance ,and compression resistance performance ,PBIFC can meet the biomechanics requirements of clinical implant .

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chongqing Medicine Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chongqing Medicine Ano de publicação: 2014 Tipo de documento: Artigo