Feasibility of magnetic resonance diffusion-weighted intravoxel incoherent motion imaging and its reproducibility in normal liver / 实用放射学杂志
Journal of Practical Radiology ; (12): 623-626, 2014.
Artigo em Chinês | WPRIM | ID: wpr-446154
ABSTRACT
Objective To investigate the feasibility of magnetic resonance diffusion-weighted intravoxel incoherent motion (DW-IVIM)imaging and its reproducibility in normal liver.Methods Thirty-five young healthy volunteers underwent liver DW-IVIM ex-amination twice on a 1 .5 Tesla Philips MR scanner.Parameter maps of D,D* and f were generated by using a manufacturer-sup-plied software.Six Region-of-interests (ROIs)(3 ROIs in left and right liver lobe,respectively)were drawn on three consecutive middle slices,respectively.The average values of D,D* and f were measured and compared between the left and right liver lobe. The reproducibility of D,D* and f measurement were assessed with Bland-Altman method.Results The values of D,D* and f of left liver lobe [D(×10-3 mm2/s)1.250±0.210,D*(×10-3 mm2/s)101.906±15.063 ,f(%)22.954±5.268]were statistically higher than those of right liver lobe[D(×10-3 mm2/s)1.039±0.124,D*(×10-3 mm2/s)100.127±20.757,f(%),18.317± 3.671](D,f,P<0.001,D*,P<0.232-0.605),however ,their reproducibility of left liver lobe [limit of agreement(LOA)22.55%,51.05%,39.3%,respectively]were lower than those of right liver lobe (LOA20.3%,39.25%,33.9%).The D values had the best reproducibility in both the left and right liver lobe.Conclusion The liver DW-IVIM is feasible.The values of D,D*and f in left liver lobe were higher that those in right liver lobe but with worse reproducibility of measurement.The D value was most reproducible in both liver lobes.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Practical Radiology Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Practical Radiology Ano de publicação: 2014 Tipo de documento: Artigo