Breakage of thoracic epidural catheter: How can we deal with it?: A report of 2 cases
Anesthesia and Pain Medicine
;
: 192-195, 2010.
Artigo
em Coreano
| WPRIM
| ID: wpr-44617
ABSTRACT
Breakage of epidural catheter during insertion and removal is a rare but dilemmatic complication. Although it is favorable to remove a broken catheter entrapped in tissue, there is a controversy about whether retained fragments should be managed even in asymptomatic patients. We experienced two cases of breakage of epidural catheter in patients with thoracic epidural analgesia. One case happened during removing of the entire catheter and needle together and the other, during the removal of the catheter alone. Because neurologic or infective complications were not detected until 14 days after the events, surgical interventions were withheld. In both cases, there were no sequelae on long-term follow up.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Cateterismo
/
Analgesia Epidural
/
Seguimentos
/
Catéteres
/
Analgesia
/
Agulhas
Tipo de estudo:
Estudo observacional
/
Estudo prognóstico
Limite:
Humanos
Idioma:
Coreano
Revista:
Anesthesia and Pain Medicine
Ano de publicação:
2010
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS