Comparison of Ocular Biometry and Postoperative Refraction in Cataract Patients Between Lenstar(R) and IOL Master(R)
Journal of the Korean Ophthalmological Society
;
: 833-838, 2012.
Artigo
em Coreano
| WPRIM
| ID: wpr-45165
ABSTRACT
PURPOSE:
To compare axial length, anterior chamber depth, and keratometric measurements of an optical low-coherence reflectometry device with those of other ocular biometry devices and evaluate the accuracy of predicting postoperative refraction.METHODS:
A total of 32 eyes in 32 patients who received cataract surgery were included in the present study. The axial length, anterior chamber depth, and keratometry were measured by optical low-coherence reflectometry (Lenstar LS900(R)), partial coherence interferometry (IOL master(R)), and ultrasound. The SRK/T formula was used to calculate IOL power, and predictive error that subtracts predictive refraction from postoperative refraction was compared among ocular biometry devices.RESULTS:
Axial length, anterior chamber depth, and keratometry had a strong correlation and demonstrated no statistically significant differences between Lenstar LS900(R) and other devices. The Bland-Altman plots showed a high degree of agreement between Lenstar LS900(R) and other devices. The mean absolute prediction errors in Lenstar LS900(R) and IOL master(R) were not significantly different.CONCLUSIONS:
The ocular biometric measurements and prediction of postoperative refraction using Lenstar LS900(R) were as accurate as IOL master(R) and ultrasound.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Catarata
/
Biometria
/
Olho
/
Interferometria
/
Câmara Anterior
Limite:
Humanos
Idioma:
Coreano
Revista:
Journal of the Korean Ophthalmological Society
Ano de publicação:
2012
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS