Verification of dose parameters under reference and non-reference conditions for radiotherapy in Liaoning province / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 456-460, 2014.
Artigo em Chinês | WPRIM | ID: wpr-453538
ABSTRACT
Objective To verify the reliability of dose parameters of radiotherapy under reference and non-reference conditions by using TLD.Methods Dose parameters were verified by using TLDs under reference and non-reference conditions,including the maximum dose in axel of 5 electron beams with energy of 9 MeV and the variations of dose by depth,source-skin distance,exposure field and 45° wedge for 10 photon beams with energy of 6 MV in 5 hospitals.Results The average relative deviation of 6 MV photon beam measured between TLDs and finger ionization chambers were 4.45%,within ± 7% as required by IAEA.The average relative deviation of 9 MeV electron beam measured between TLDs and plane parallel chambers were 2.45%,within ± 5% was required by IAEA.Conclusions Measuring dosimetric parameters by using TLDs under reference and non-reference conditions was reliable and feasible.