Fully Automated Determination of Four Androgenic Hormones in Serum by Online Turbulent Flow Solid Phase Extraction Coupled with Liquid Chromatography-Tandem Mass Spectrometry / 分析化学
Article em Zh | WPRIM | ID: wpr-458386
Biblioteca responsável: WPRO
ABSTRACT
A novel method was developed for the direct analysis of testosterone, androstenedione, methyltestosterone and methenolone in serum samples by fully automated online turbulent flow solid phase extraction coupled with high performance liquid chromatography-tandem mass spectrometry. An aliquot of 50 μL serum sample was preconcentrated directly on a Turboflow SPE column after centrifugation. Turboflow SPE C18-P could be used to remove serum matrix effectively. The optimum loading flow rate and elution time were 4. 0 mL/min and 1. 0 min, respectively. The linearity ranges were from 1. 0 μg/L to 100. 0 μg/L for four target compounds. The method limits of detection ( LODs) were in the range of 0. 2-0. 3 μg/L. The relative standard deviations (RSDs) ranged from 2. 9% to 14. 1% (n=5). The time for one sample analysis including extraction, separation and determination was 32 min. The proposed method has been successfully applied for the analysis of serum samples.
Palavras-chave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: Chinese Journal of Analytical Chemistry Ano de publicação: 2014 Tipo de documento: Article