Your browser doesn't support javascript.
loading
The Usefulness of the PCL Power Calculation Computer Program: 'POWER!'
Journal of the Korean Ophthalmological Society ; : 23-28, 2002.
Artigo em Coreano | WPRIM | ID: wpr-45864
ABSTRACT

PURPOSE:

To assess the usefulness and reliability of the IOL power calculation computer program ('POWER!').

METHODS:

Calculation times and projected postoperative refractive values of 100 patients with 'POWER!' program were compared with those of the program mounted on the Humphrey R A/B scan system (model 820) by SRK-II and SRK/T formula.

RESULTS:

There was no significant difference between the projected postoperative refractive values of 'POWER!' program and Humphrey R A/B scan system (paired-t test, p>0.05). Calculation time of 'POWER!' was shorter than that of the program mounted on the Humphrey R A/B scan system incalculating more than 3 sets of IOL (paired-t test, P0.05).

CONCLUSIONS:

'POWER!' is a useful and reliable program for IOL power calculation using SRK-II and SRK/T formula.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Limite: Humanos Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Ano de publicação: 2002 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Limite: Humanos Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Ano de publicação: 2002 Tipo de documento: Artigo