Comparative Study on Concentration Monitoring of CsA in Human Whole Blood by EMIT and HPLC / 中国药师
China Pharmacist
;
(12): 2003-2006, 2014.
Artigo
em Chinês
| WPRIM
| ID: wpr-458795
ABSTRACT
Objective:
To compare the difference and correlation of HPLC and enzyme-multiplied immunoassay test( EMIT) for the determination of CsA in human whole blood.Methods:
A total of 119 clinical samples at different concentrations of CsA were collected and respectively determined by HPLC and EMIT. The difference and correlation of the two determination methods were investigated.Results:
There was significant difference in the blood concentrations of CsA determined by HPLC and EMIT(P<0. 05). CsA concen-tration determined by EMIT was 26. 2 ng·ml-1 higher than that determined by HPLC, and 95% CI was (14. 6-37. 7) ng·ml-1 . A satisfactory correlation was achieved between the two methods(r=0. 997 4).Conclusion:
There is statistically significant difference in the CsA concentration in whole blood respectively determined by EMIT and HPLC. Attention should be paid to CsA monitoring by E-MIT and HPLC, and relevant adjustment should be carried out.